Computer Interfacing
Teacher Information: The Effect of Alcohol on Biological Membranes
Suggestions and Tips
- The light source for the red-pigmented solution is the blue
LED (470 nm).
- Isopropyl alcohol may be used in place of 1-propanol, if necessary.
If rubbing alcohol is used, remember that it is already diluted
to 70% or so. The actual dilution is usually listed on the jar.
You will need to recalculate the volumes used for this alcohol.
- Use 95% ethanol and absolute methanol.
- Beets will stain hands and clothing. Lab aprons and gloves
should be worn to prevent staining.
- Obtain large, red beets from a local store. Students will
cut these into small 1 cm X 1 cm X
0.5 cm pieces. To save class time, you may want to cut
a 1 cm X 1 cm rectangle for each group. They
can slice 0.5-cm sections from this. The end pieces should be
discarded, however, as they will dry out and behave differently
than freshly cut surfaces.
- The cuvette must have at least 1.5 mL of solution in order
to get a valid absorbance reading. If students fill the cuvette
as described in the procedure, they should easily be in this range.
- Since there is some variation in the amount of light absorbed
by the cuvette if it is rotated 180°, you should use a water-proof
marker to make a reference mark on the top edge of one of the
clear sides of all cuvettes. Students are reminded in the procedure
to align this mark with the white reference mark to the right
of the cuvette slot on the colorimeter.
- The use of a single cuvette in the procedure is to eliminate
errors introduced by slight variations in the absorbance of different
plastic cuvettes. If one cuvette is used throughout the experiment
by a student group, this variable is eliminated.
- The use of cotton swabs is recommend to dry a cuvette after
water or a solution has been emptied from it. The cotton swab
will remove any remaining droplets in the cuvette. After completely
drying the cuvette using both ends of the cotton swab, it is no
longer necessary to rinse with the new solution.
- Satisfactory results can be obtained if a different cuvette
is used for each solution of the experiment. For best results,
the cuvettes for one student lab team should be matched.
Each cuvette in a matched set absorbs light (when empty) at approximately
the same level. To match a set of cuvettes, first calibrate the
colorimeter using the method described in Appendix A. Use a clean,
dry cuvette for the 100% calibration instead of a distilled water
blank. Put a reference mark on one of the clear sides of the cuvette
so it is always oriented the same way in the cuvette slot. Place
each cuvette in the batch into the colorimeter and record percentage
transmittance values for each. When you are finished, group the
cuvettes according to similar %T values. Each of these groups
represents a set of matched cuvettes.
- This experiment gives you a good opportunity to discuss the
relationship between percentage transmittance and absorbance.
At the end of the experiment, students can press the mouse on
the absorbance label of the vertical axis to access the pop-up
menu showing the columns. If they choose the column "T"
(percentage transmittance), they will change the graph to a plot
of percent transmittance vs. concentration. You can also discuss
the mathematical relationship between absorbance and percent transmittance,
as represented by either of these formulas:
A = log(100 / %T) or A = 2 - log(%T)
|
|
|